3.8.2.1. Continuous Mode and a Storage Qualifier Examples
The following show the capture differences between a non-segmented buffer in continuous mode and a non-segmented buffer using a storage qualifier. The configuration of the logic analyzer waveforms is a base trigger condition, sample depth of 64 bits, and Post trigger position.
In the continuous data capture, Trig1 occurs several times in the data buffer before the Signal Tap logic analyzer trigger activates. The buffer must be full before the logic analyzer evaluates any trigger condition. After the trigger condition occurs, the logic analyzer continues acquisition for eight additional samples (12% of the buffer, as defined by the "post-trigger" position).
In conditional capture the logic analyzer triggers immediately. As in continuous capture, the logic analyzer completes the acquisition with eight samples, or 12% of 64, the sample capacity of the acquisition buffer.